Order number 1808-0506 -

-
Get a quote
ContactOnline shop
BioShake.com
-
Overview
Small . Light . Efficient . Quit
Compact and stylish designed, the BioShake iQ high-speed thermoshaker lets you perform all your standard runs with a minimum of adjustments, and offers outstanding performance to handle a wide range of applications across biotechnology, pharmaceutical and academic research.
By combining the mixing operation with the incubation phase, reaction process times and operator workload are reduced and efficiency of many procedures is increased, resulting in a higher throughput.
Wide range of applications
The BioShake iQ comes with a variety of standardized and specific adapter plates. The adapters allow an optimal fit for standard tubes, lysis tubes, microplates, glass vials and other sample vessels. An excellent temperature uniformity and homogeneity is guaranteed by using these adapters.
Gentle mixing of samples by planar orbital motion
German designed and manufactured, BioShake iQ thermoshaker offer an ultra-efficient, 2-dimensional shaking axis so that samples mix completely in a fraction of the time of competing systems. The mixing orbit of 2.0 mm is always constant. The orbital shaking is precisely controlled, in fact, that you need never spin down your plates after mixing. Even tubes, vials, high density plates, or low sample volumes, offer no obstacle for these precision tools.
Fully adjustable between 200 rpm and 3,000 rpm, well beyond the speeds of most other brands, guarantees fast, splatter-free, mixing for tubes, glass vials or across an entire 384-well microplate.
Homogeneous and accurate temperature control
Within the BioShake iQ thinnest, large-area heating elements and sensors are integrated. The heat output of 75 W can be adjusted seamlessly from room temperature to 99°C. The equipment is fitted with a new and very reliable electronic temperature control which ensuring optimum warming of samples. During the heat up time the special electronic control system guarantees the same temperature for all vessels with minimal deviations. Precisely machined adapter plates allow an individual adaptation to the used sample holder.
Stylish aluminium housing
The first-class finished aluminium housing gives the BioShake iQ its essential functionality. It provides a high amount of security, device stability and ensures a long service life.
Discover why BioShake iQ is without peer in the industry.
-
Details
-
Safety

Safety of the user and ease of use are clearly a priority for us. Independent investigations from TÜV services and CE certifications guarantee the highest security standards.
The quality label "Made in Germany" means to you:
- Years of trouble free operation.
- The sealed housing to protect mechanical and electronic components increases the lifespan of 10 years on average of well maintained equipment.
- All units are designed for continuous 24 hour unattended operation when utilizing sound scientific methods.
- To prevent laboratory fires, all units feature an over temperature circuit which switches off if an over temperature situation occurs.
- Thermal damage to any unit is minimized or prevented, as all models come with a fire resistant aluminum housing.
Temperature control
The BioShake iQ heats from ambient up to 99 °C in a short time, provides ease of use features, amazing comfort and maximum safety.
Within the BioShake iQ thinnest large-area heating elements and sensors are integrated. During the heat up time the special electronic control system guarantees the same temperature for all vessels with minimal deviations.
Temperatures are selectable from ambient temperature to 99°C in steps of 1.0°C.
The nominal regulation accuracy is better than ±0.1°C with a uniformity of temperature distribution better than ±0.5°C at 45°C across the heating surface.Exchangeable adapter plates and thermal blocks enable a perfect adaptation to microplates, tubes or glass vials.
Comfort
Programming the BioShake iQ works via direct touch buttons. In addition two buttons for start and storage of time and mixing modes enable the instrument to run complex applications. This opens new points of view on the daily laboratory work and optimizes routine application enormously.
The short mix button allows short and fast mixing in between.
The two line LCD display guarantees simultaneous and safe reading of all programmed and measured parameters as time, mixing frequency and for BioShake iQ additional temperature.
-
-
Knowledge
-
Optimization of mixing results
The adjustment of the optimal mixing frequency for microplate or tubes should always be made in dependence on the size of the well and the filling volume.
Only in this way optimum results can be achieved with in shortest process time with a high reproducibility.Recommended mixing speeds for microplates
Mixing speeds [rpm] for different microplates depending on filling volume/well [%] *
Filling volume
in %96 well
(standard)384 well
(standard)384 well
(small-vol.)1536 well
(standard)10 %
1800 - 2200
2200 - 2600
2800 - 3000
4000 - 5000
25 %
1600 - 2000
2000 - 2400
2400 - 3000
3500 - 4500
50 %
1400 - 1800
1800 - 2200
2200 - 2600
2800 - 3500
75 %
1200 - 1600
1600 - 2000
2000 - 2400
2500 - 3000
* All values refer to diluted fluids and mixing orbit of 2.0 mm.
Values for 1536 well plates refer to mixing orbit of 1.2 mm.Recommended mixing speeds for tubes & vials
Mixing speeds [rpm] for tubes/vials depending on filling volume [%] *
Filling volume
in %0.2 ml tube
0.5 ml tube
1.5 ml tube
2.0 ml tube
10-50 %
1400 - 1800
1200 - 1600
1000 - 1300
1000 - 1300
50-75 %
1200 - 1500
1100 - 1300
1000 - 1200
900 - 1200
75-100 %
1000 - 1300
1000 - 1200
900 - 1100
900 - 1100
* All values refer to diluted fluids and mixing orbit of 2.0 mm.
Application notes . Lab reports . Publications . Presentations . Case studies . Literature
Physics of mixing - "A Methodology and Study of Microplate Mixing Techniques"
Optimization of mixing parameters in lab automation for high & low sample volumes
Mixing efficiency of Bioshake 3000 plate shaker - Authored by: Tanya Knaide, Artel, Inc.
Experimental studies of mixing efficiency in 96 well microplates
Experimental studies of mixing efficiency in standard tubes
Videos for optimization of mixing frequencies
Optimization of temperature control in microplates
-
-
Specification
Adapter Plates
Microplates
96-, 384-, and 1536-well microplates, deep well plates, PCR plates
(e.g Eppendorf®, Brandt®, Nunc®, Greiner Bio-One®)Tubes
0.2 / 0.5 / 1.5 / 2.0 ml standard microcentrifuge tubes
(e.g. Eppendorf®: Standard tubes, Safe-Lock microcentrifuge tubes, PCR tubes)Glass vials
2.0 and 4.0 ml glass vials
(e.g. Zinsser-Analytic®)others
on request
Temperature
Temperature range
ambient to 99°C
Temperature setting
0.1°C increment
adjustable from 0°C to 99°CTemperature accuracy
±0.1°C
Temperature uniformity
±0.5°C at 45°C
±0.7°C at 75°C
±1.0°C at 95°CHeat-up time
ca. 7°C/min
(from ambient to 95°C in 10 min)Mixing
Mixing frequency
200 to 3,000 rpm (microplates)
200 to 1,800 rpm (tubes, glass vials)Mixing orbit
constant 2 mm diameter
Speed setting resolution
50 rpm increments
Mixing regulation accuracy
± 25 rpm
Short-Mix function
Yes
Timer
Timer setting
1 min - 99 h
with automatic switch to stand-byTimer setting resolution
1 min
Readability
Minutes, Seconds
Continuous working
Yes
Audible Alarm
Yes
Programming
Programs stored
2
Definable buttons
P1 and P2
Individual program capacity
3 steps
Internal memory
Yes
Display
Display
LCD-display with backlight (blue)
Target values
Time, Mixing frequency, Temperature
Actual values
Time, Mixing frequency, Temperature
Electrical
Controller
Micro controller
Power switch
Yes
Operating Voltages
24 V DC input
100 WattPower supply
External power supply unit
100-240 V AC, 50-60 HzProperties
Housing Material
Aluminum anodized
Environment operating range
+5°C to 45°C (80 % max. relative humidity)
Dimensions (W x D x H)
142 mm x 170 mm x 80 mm
(5.59 in x 6.69 in x 3.14 inches)Weight
2.7 kg (5.95 lbs)
-
Accessories
-
-
Changeable adapter plates
For all BioShake units Q.Instruments offers high precision adapter plates to allow a perfect fit for all kinds of standard tubes, vials, microplates and other different disposables. The adapter plates are optimized for an excellent heat transfer to the disposables and enhance the uniformity over all wells and the heat up or cool down time. The exchange of adapter plates can be performed very easily within one minute.
-
Microplates
Standard
1808-1021 | Microplate . Flat bottom standard
1808-1022 | Microplate . Flat bottom High Base
1808-1024 | Microplate . Flat bottom Low Base
1808-1031 | Microplate . 96 well round type 1
1808-1032 | Microplate . 96 Well round type 2
PCR microplates
1808-1041 | PCR Plate . 96 well
1808-1051 | PCR Plate . 384 well
Deep well plates
1808-1121 | Deep well . Eppendorf® 96/1000
1808-1131 | Deep well . Eppendorf® 96/500
1808-1141 | Deep well . BRAND® 96/1100
1808-1151 | Deep well . NUNC® 96/2000 -
Tubes & vials
Conical shape
1808-1061 | Tubes . 24x 2.0 ml . 15x 0.5 ml
1808-1062 | Tubes . 24x 1.5 ml . 15x 0.5 ml
1808-1063 | Tubes . 40x 0.5 ml . 28x 0.2 ml
1808-1064 | Tubes . 96x 0.2 ml
1808-1065 | Tubes . 35x 2.0 ml . 24x 0.5 ml
1808-1066 | Tubes . 35x 1.5 ml . 24x 0.5 ml
Cylindrical shape
1808-1067 | Lysis vials . 35x 0.5-2.0 ml
1808-1069 | Vials . 35x 2.0 ml Ø 10.8 mm
1808-1071 | Vials . 30x 2.0 ml Ø 12 mm
1808-1072 | Vials . 20x 4.0 ml Ø 15 mm
1808-1073 | Vials . 20x 4.0 ml Ø 17 mm
1808-1074 | Vials . 20x 6.0 ml Ø 19 mm -
Special versions
1808-1081 | 5x Alere® ArrayStrip
1808-1085 | 24x Alere® ArrayTube
For detailed information please click on the particular 8-digit order number.
Graphical overview

Special adapter and customized solutions

You have not found a suitable adapter for your BioShake?
We develop and design your adapter according to your sample carrier and specification as follows.
Time frame
Proceeding
Day 1




Please send us 1 sample and specific information about your used sample carrier: manufacturer, description, article number.
Day 2-6
We will develop and design the adapter according to your specification.
Day 7
We will send you a formal confirmation letter with solution description.
Day 7-10
You will check the proposed solution and we carry out any adjustments.
Day 11-30
After receiving the approval of design we will start with the production.
Form to specify your application and requirements
Please use the
S15 service form for customized adapters to specify your application and requirements.
An accepted specification needs to contain the following complete information: name of the sample container, name of the manufacturer, article number, and a general description of the area of application.
Please fill in this form on page 2 and send it to us by email, mail or fax.
-
Spare parts & service accessories for BioShake
-
For all BioShake units and automation modules, various spare parts, mounting and service accessories are always available.
Order number Name Mounting accessories 1808-9008 Security cover plate for BioShake ff. 1808-9012 Allen key for adjusting microplates (BioShake 3000, BioShake 3000-T) 1808-9013 Grounding set for BioShake 3000 ff. 1808-9015 Anti slide pad for BioShake 3000 ff. 1808-9016 Bumpers for BioShake ff. (set of 4 pieces) Communication 1808-9120 USB/RS232 Converter (Digitus USB serial adapter USB 2.0) 1810-0061 Daisy Chain Box 4 (for connection of 2-4 BioShake units) 1810-0065 Daisy Chain Box 8 (for connection of 5-8 BioShake units) Power supply & cords 1808-9011 External power supply unit (100-240 V AC, 50-60 Hz, 24 V DC, 120 W) 1808-9101 Power cord with 3-pin plug (clover leaf) - Europe 1808-9102 Power cord with 3-pin plug (clover leaf) - Switzerland 1808-9103 Power cord with 3-pin plug (clover leaf) - UK 1808-9104 Power cord with 3-pin plug (clover leaf) - Italy 1808-9110 Power cord with 3-pin plug (clover leaf) - USA 1808-9111 Power cord with 3-pin plug (clover leaf) - Japan 1808-9112 Power cord with 3-pin plug (clover leaf) - China 1808-9113 Power cord with 3-pin plug (clover leaf) - Korea 1808-9114 Power cord with 3-pin plug (clover leaf) - Australia Service accessories 1808-9001 Service set of axial bearing balls for BioShake (set of 3 pieces) 1808-9002 Service set of O-rings coated (special slide way coating, 13x2mm, set of 3 pieces) 1808-9003 Service Pack for BioShake (complete set of spare parts) 1808-9004 Service set of ELM-Pins for BioShake 3000 elm (set of 4 pieces) 1808-9005 Service set of mass balance elements for BioShake 1808-9006 Service set of fixing screws for BioShake adapter plates 1808-9007 Service set of clamping springs BioShake XP/iQ 1808-9009 Service set of ELM-Pins for BioShake 3000-T elm (set of 4 pieces) 1808-9014 Sevice set of ELM-Springs for BioShake (set of 2 pieces)
-
-
-
Order details
-
Scope of delivery
- 1x BioShake iQ
- 1x Power supply 110-240 VAC, 24 VDC
- 1x Power cord - Europe
- 1x Power cord - Country specific version
- 1x Calibration certificate
- 1x Documentation
Adapters are not included in delivery and have to be ordered separately - see accessories.
-
2 years global warranty
Q.Instruments warranted for the BioShake models an extended warranty of 2 years - as an expression of their superior quality in terms of functionality, performance, workmanship, durability and safety.
For a faster customer service, we recommend a free online device registration after purchasing.
Online Device RegistrationYou need a quotation?
Please contact us via online
contact form or directly to:
QUANTIFOIL Instruments GmbH
Loebstedter Straße 101
07749 Jena
Germany
Phone: +49 3641 87612-0
Fax: +49 3641 87612-99
E-Mail:
info[@]QInstruments.comDo you want to order online?
Please visit our
Online Shop at www.BioShake.com.
Carefree purchase orders and invoices
Customers can pay by invoice. We recommend prepayment if you’re ordering from abroad.
Payment methods
You can choose from the following payment options:
1. Wire transfer
2. Direct debit
3. Payment via check
4. Payment via credit card: Visa, MasterCard or JCB
5. Payment via Paypal
more information...
-
-
Downloads
-
-
Product flyer / Catalog

Product flyer
559 K 
1808-xxxx_BioShake_adapter_v001.pdf
Changeable thermo adapter plates for BioShake
626 K Manuals

Operating Manual
1.9 M 
shortmanual_bioshake_iq_en.pdf
Short Manual
1.4 M 
sdoc_iq_xp_adapter_changing_en_v001.pdf
Manual for changing adapters
763 K 
scard_bioshake_blocks_en_de.pdf
Short manual for changing adapters
675 K Applications

poster_slas2012_microplate_mixing.pdf
Poster - Optimization of mixing parameters in lab automation for high & low sample volumes
0.9 M 
poster_labauto2011_optimix.pdf
Poster - Optimization of mixing processes in microplates
2.3 M 
BioShake3000_MixingEfficiency_ApplNote.pdf
BioShake - Mixing Efficiency - Appl. Note 1 - Authored by Tanya Knaide, Artel, Inc.
63 K
-
-