FAQ - las preguntas más frecuentes
Here you can find answers for frequently asked questions asked by our users.
-
Which tubes and sample containers can be used in the BioShake XP/iQ?
-
Depending on the type of exchangeable thermoblock being used the BioShake can accommodate the following kind of tubes and sample containers:
- 96 x 0.2 ml tubes
- 40 x 0.5 ml tubes
- 24 x 1.5 ml tubes
- 24 x 2.0 ml tubes
- 30 x 2.0 ml glass vials 12 mm diameter
- 20 x 4.0 ml glass vials 15 mm diameter
- 24 x 1.5 ml Alere ArrayTubes
- 5 x Alere ArrayStrip (5 stripes)
- 12 x Alere ArrayStrip (12 stripes)
- 1 x Microplate (universal)
- 1x 96, 384 or 1536 micro well plates
- 1x PCR-plate
- 1x Deep well plate
-
What is the SBS Standard for Microplates?
-
The Society for Biomolecular Sciences (SBS; www.sbsonline.org/msdc) collaborates with other institutions to develop standards for micro test plates – the objective is to standard the various formats for easier automation.
For use with BioShake the footprint of microplates should meet the following specifications:
- Rectangular plate
- Length = 127.76 mm +/- 0.5 mm
- Width = 85.48 mm +/- 0.5 mm
- The footprint must be continuos and uninterrupted around the base of the plate.
-
Which international standards and specifications are comply with BioShake devices?
-
All BioShake devices are conform to EU requirements, all latest safety regulations and DIN standards (State of 2010/06/01):
- Elektromagnetische Verträglichkeit 2004/108/EG
- Niederspannungsrichtlinie 2006/95/EG
- DIN EN 61010-1 :2002-08
- DIN EN 61010-2-010 :2004-06
- DIN EN 61010-2-051 :2004-07
-
Which microplates can be used on the BioShake?
-
All plate formats that match to the SBS Standard for microplates, for example:
- MTP (up to 1536-Well)
- 96- and 384-well PCR-plates (skirted, semi-skirted, unskirted)
- Deep well plates (up to 384-Well)
- Cell culture plates (6-, 12-, 24-, 48-, 96- and 384-well)